FDA: LifeVest wearable defibrillator has safety issue


 

The Zoll LifeVest 4000, a wearable defibrillator, could fail to deliver a treatment shock after displaying the message “Call for service: Device has a problem that may require service. Call ZOLL for service, Message Code 102,” according to the FDA.

“Failure to contact Zoll and immediately replace the device after Message Code 102 appears on the device screen may result in serious patient harm or death of the patient because the device may fail to deliver therapy appropriately when needed” according to an FDA press release.

FDA icon
When functioning properly, the LifeVest is intended to deliver a shock to correct a patient’s heartbeat to a normal rhythm. However, when the error mentioned above occurs, LifeVest may not be able to function as intended and deliver the corrective shock because of an issue that prevents the device from charging its high-energy capacitors. To compound the malfunction of the device, the error message “Message Code 102” does not indicate that the device is not functioning properly and that the patient needs to contact Zoll immediately.

Only one death associated with “Message Code 102” malfunction of LifeVest has been reported, but about 0.1% of devices have displayed the “Message Code 102” error. According to Zoll, roughly 33,670 devices have been distributed as of Nov. 14, 2017, with nearly 75% of them distributed in the United States.

On Jan. 14, Zoll issued a voluntary recall of the LifeVest 4000. The FDA has indicated that it will continue to work with Zoll to monitor adverse events associated with the “Message Code 102” error and work on finding a permanent solution to this problem. Recommendations for physicians, caregivers, and patients regarding how to respond to error messages can be found here.

Recommended Reading

Losmapimod failed to beat placebo in acute MI trial
MDedge Emergency Medicine
Early antiarrhythmic drugs boost survival in shock-refractory cardiac arrest
MDedge Emergency Medicine
Cardiologists underrecognize angina about half the time
MDedge Emergency Medicine
Medical quality beats patient comfort
MDedge Emergency Medicine
New trials address unresolved issues in endovascular therapy for acute ischemic stroke
MDedge Emergency Medicine
CSL112 enhances cholesterol efflux capacity after acute MI
MDedge Emergency Medicine
Prenotification, unequivocal stroke promote ultra-fast door-to-needle time
MDedge Emergency Medicine
Forgo supplemental oxygen in adequately perfused patients with acute MI, study suggests
MDedge Emergency Medicine
U.S. judge orders Philips to cease AED manufacturing
MDedge Emergency Medicine
VIDEO: Regionalized STEMI care slashes in-hospital mortality
MDedge Emergency Medicine