Original Research

Revision of Hip Resurfacing Arthroplasty

Author and Disclosure Information

Metal-on-metal (MOM) hip resurfacing has become an increasingly popular treatment for young, active patients with degenerative disease of the hip, as bearing surfaces with better wear properties are now available. One proposed advantage of resurfacing is its ability to be successfully revised to total hip arthroplasty (THA). In addition, radiographic parameters that may predict failure in hip resurfacing have yet to be clearly defined.

Seven MOM resurfacing arthroplasties were converted to conventional THAs because of aseptic failure. Using Harris Hip Scores (HHS) and Short Form 12 (SF-12) questionnaire scores, we compared the clinical outcomes of these patients with those of patients who underwent uncomplicated MOM hip resurfacing. In addition, all revisions were radiographically evaluated. Mean follow-up periods were 51 months (revision group) and 43 months (control group).

There was no significant difference between the 2 groups’ HHS or SF-12 scores. There was no dislocation or aseptic loosening after conversion of any resurfacing arthroplasty. Valgus neck-shaft angle (P<.03) was associated with aseptic failure of MOM hip resurfacing.

Conversion of aseptic failure of hip resurfacing to conventional THA leads to similar clinical outcomes for patients who undergo uncomplicated hip resurfacing. The orientation of the femur and the components placed play a large role in implant survival in hip resurfacing. More work needs to be done to further elucidate these radiographic parameters.


 

Recommended Reading

Resurfacing of Isolated Articular Cartilage Defects in the Glenohumeral Joint With Microfracture: A Surgical Technique & Case Report
MDedge Surgery
Incidence of Early Development of Radiolucent Lines in Keeled Polyethylene Glenoid Components After Total Shoulder Arthroplasty
MDedge Surgery
Arthroscopic Management of a Chronic Primary Anterior Shoulder Dislocation
MDedge Surgery
New Products/Product News
MDedge Surgery
New Products/Product News
MDedge Surgery
New Products/Product News
MDedge Surgery
Clinical Presentation of Pulmonary Embolus After Total Joint Arthroplasty: Do Size and Location of Embolus Matter?
MDedge Surgery
Exostosis-Like Intra-articular Periosteal Osteoblastoma: A Rare Case
MDedge Surgery
Unusual Presentation of Cobalt Hypersensitivity in a Patient With a Metal-on-Metal Bearing in Total Hip Arthroplasty
MDedge Surgery
Failed Metal-on-Metal Total Hip Arthroplasty Presenting as Painful Groin Mass With Associated Weight Loss and Night Sweats
MDedge Surgery