SVS: Five-year data support Endurant stent durability

View on the News

Long-term EVAR outcomes as good as open repair

What’s refreshing with this particular trial is that they showed us what happens in 5 years. There’s no question that EVAR has lower mortality than open repair immediately postop, but there have been arguments whether long-term survival is as good. The answer is ‘yes.’ The thing that impressed me the most was the freedom from aneurysm-related mortality at 99%. That’s phenomenal. It means that this device is very solid and very durable. At 5 years, they had a few type II endoleaks. I don’t think that’s too high compared to other trials. We are getting into the era when [these grafts] are very comparable. Selection comes down to anatomy, what physicians are comfortable with, and, to be honest, representative availability.

Dr. Mahmoud Malas is chief of endovascular surgery at Johns Hopkins Bayview Medical Center in Baltimore. He said he has no financial or other relationships with Medtronic.


 

AT THE 2015 VASCULAR ANNUAL MEETING

References

CHICAGO – During 5 years of follow-up, there was one aneurysm-related death among 150 patients who received Endurant stent grafts for abdominal aortic aneurysms, according to outcome data from Endurant’s manufacturer, Medtronic.

The single death was in a man who developed a type I endoleak 2 years after implant. He declined further intervention, presented at about 3 years with a ruptured aneurysm, and died shortly thereafter. If that patient had been treated, it’s possible that Endurant’s freedom from aneurysm-related death would have been 100% at 5 years, instead of 99%.

Dr. Michael Singh

Dr. Michael Singh

Endurant “has proven to be safe and effective through 5 years of follow-up. Long-term outcomes indicate that it’s a durable repair with very low aneurysm-related mortality and a limited need for secondary interventions,” said Dr. Michael Singh, associate professor of surgery at the University of Pittsburgh, who presented the findings at a meeting hosted by the Society for Vascular Surgery.

The presentation completed Medtronic’s U.S. regulatory trial. The Food and Drug Administration approved the device in 2010 based on 30-day safety and 12-month efficacy data. Endurant is currently used in about half of endovascular triple A repairs worldwide, Medtronic said in a press release announcing the 5 year results.

The 150 subjects in the trial had aneurysms of at least 50 mm, with neck lengths of at least 10 mm, neck angulations no more than 60 degrees, and iliac fixation lengths of at least 15 mm. They were 73 years old, on average, and most were men. Endurant deployment was successful in all but one.

Twenty-five patients (17.7%) died during follow-up, all but the type I endoleak patient from causes expected in a triple A cohort, including stroke and lung disease. Six additional patients were lost to follow-up, and 18 withdrew for a variety of reasons, including physician advice.

At 5 years, aneurysm sacks had shrunk in about two-thirds of patients, and remained stable in most of the rest. Endurant didn’t fracture, migrate, or disconnect during follow-up.

Fifteen patients required 18 secondary endovascular interventions, including two for type I endoleaks, 10 for type II endoleaks, three for limb occlusion before 2 months, one for limb occlusion after 2 months, one for limb stenosis, and one for thromboembolism.

When asked to explain the robust outcomes, Dr. Singh noted that “we are dealing with trial patients ... with very selective inclusion and exclusion criteria” and favorable anatomy, “so you expect optimal results. In real-world experience, we might see something slightly” less optimal.

The type I endoleak patient who declined treatment might have had “the device placed in a degenerated aorta. We have come across this in our own practice regardless of the device that is utilized. I think the foundation” of endograft placement “is making sure that you are deploying the device in normal aorta,” at least as much as possible, he said.

The trial was funded by Medtronic. Dr. Singh said he has no relevant disclosures. The principal investigator is a Medtronic consultant.

aotto@frontlinemedcom.com

Recommended Reading

Elephant stent aorta repair – good outcomes, but is it too complex?
MDedge Cardiology
Post-CABG stroke risk same with one or two clamps
MDedge Cardiology
Monitoring effectively identifies seizures in postbypass neonates
MDedge Cardiology
Idarucizumab reverses dabigatran’s anticoagulant effects
MDedge Cardiology
Two U.S. transcatheter valve approvals reshape TAVR
MDedge Cardiology
Older MI patients missing out on ICDs
MDedge Cardiology
ACC/HRS/SCAI seek to unify rollout of percutaneous LAA occlusion
MDedge Cardiology
SVS: Four easy preop variables predict mortality in ruptured AAAs
MDedge Cardiology
Reports of TAVI leaflet thickening downplayed – for now
MDedge Cardiology
Real-time detection of aortic regurgitation during TAVI without TEE
MDedge Cardiology